Detail description

Advantages

The ultrasonic flow meter is designed to measure the fluid velocity of liquid within a closed conduit. The transducers are a non-contacting, clamp-on type, which will provide benefits of non-fouling operation and easy installation.
The transit-time flow meter utilizes two transducers that function as both ultrasonic transmitters and receivers. The transducers are clamped on the outside of a closed pipe at a specific distance from each other. The transducers can be mounted in V-method where the sound transverses the pipe twice, or W-method where the sound transverses the pipe four times, or in Z-method where the transducers are mounted on opposite sides of the pipe and the sound crosses the pipe once. This selection of the mounting method depends on pipe and liquid characteristics. The flow meter operates by alternately transmitting and receiving a frequency modulated burst of sound energy between the two transducers and measuring the transit time that it takes for sound to travel between the two transducers.

Basic parameters

ItemsSpecifications
Main unitAccuracyBetter than +1%
RepeatabilityBetter than 0.2%
PrincipleTransit-time measuring principle
Measurement Period500ms
DisplayLCD with backlight, display accumulated flow/heat, instantaneous flow/heat, velocity, time etc.
OutputAnalogue output: 4-20mA or 0-20mA current output. Impedance 0~1kΩ. Accuracy 0.1%
OCT output: Frequency signal (1~9999HZ)
Relay output over 20 source signal (no signal, reverse flow etc.)
RS485 serial port
InputThree analogue input
Three-wire PT100 resistor input (optional)
Other functionsAutomatically record the totaliser data of the last 64 days / 64 months /5 years;
The power-on time and corresponding flow rate of the last 64 power on and off events. Allow manual or automatic flow loss compensation
The instrument working status of the last 64 days
PipematerialSteel, stainless steel, cast iron, cement pipe, copper, PVC, aluminum, FRP etc. Liner is allowed
Size15-6000mm
Straight pipe sectionIn the upstream it must be beyond 10D, in the downstream it must be beyond 5D, in the upstream the length must be beyond 30D from the access of the pump. (D stands for pipe diameter)
LiquidTypesWater, sea water, industrial sewage, acid & alkali liquid, alcohol beer, all kinds of oils which can transmit ultrasonic single uniform liquid
TemperatureStandard: -30°C – 90°C , High-temperature: -30°C – 160°C
TurbidityLess than 10000ppm, with a little bubble
Flow DirectionBi-directional measuring, net flow/heat measuring
EnvironmentTemperatureMain Unit: -30°C – 80°C
Transducer: -40°C -110°C, Temperature transducer: select on enquiry
HumidityMain Unit: 85% RH
Transducer: water-immersible, water depth less than 3m
CableTwisted Pair Line, standard length of 20m, can be extended 500m (not recommended); Contact the manufacturer for longer cable requirement.
RS-485 interface, transmission distance up to 1000m
Power SupplyAC220V or DC24V
Power ConsumptionLess than 1.5W
ProtocolsMODBUS, M-BUS, Fuji extended protocol and other factory protocol